DocumentCode :
1144513
Title :
High-Speed Spiral-Scan Echo Planar NMR Imaging-I
Author :
Ahn, C.B. ; Kim, J.H. ; Cho, Z.H.
Volume :
5
Issue :
1
fYear :
1986
fDate :
3/1/1986 12:00:00 AM
Firstpage :
2
Lastpage :
7
Abstract :
An improved echo planar high-speed imaging technique using spiral scan is presented and experimental advantages are discussed. This proposed spiral-scan echo planar imaging (SEPI) technique employs two linearly increasing sinusoidal gradient fields, which results in a spiral trajectory in the spatial frequency domain (k-domain) that covers the entire frequency domain uniformly. The advantages of the method are: 1) circularly symmetric T2 weighting, resulting in a circularly symmetric point spread function in the image domain; 2) elimination of discontinuities in gradient waveforms which in turn will reduce initial transient as well as steady-state distortions; and 3) effective rapid spiral-scan from dc to high frequency in a continuous fashion, which ensures multiple pulsing with interlacing for further resolution improvement without T2 decay image degradation. Some preliminary experimental results will be presented and further possible improvements suggested.
Keywords :
Frequency domain analysis; High-resolution imaging; Image reconstruction; Image resolution; Magnetic resonance imaging; Nuclear magnetic resonance; Optical imaging; Sampling methods; Spirals; Student members;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/TMI.1986.4307732
Filename :
4307732
Link To Document :
بازگشت