• DocumentCode
    1144949
  • Title

    Time-Coded Aperture Tomography: Experimental Results

  • Author

    Van Giessen, John W. ; Viergever, Max A. ; De Graaf, Cornelis N. ; Dane, Huib J.

  • Volume
    5
  • Issue
    4
  • fYear
    1986
  • Firstpage
    222
  • Lastpage
    228
  • Abstract
    In this paper, we discuss the applicability of a time-coded aperture system??especially designed for thyroid tomography??on the basis of phantom experiments. Our studies show that 1) the quality of the reconstructions is high (e.g., a cold spot of 6 mm diameter in a thyroid phantom can easily be detected), and 2) the reconstruction can be carried out in less than 11 min on a standard 16 bit minicomputer (HP1000). It is therefore concluded that the clinical potentiality of the device is good.
  • Keywords
    Apertures; Collimators; Data acquisition; Decoding; Detectors; Hardware; Image reconstruction; Imaging phantoms; Microcomputers; Tomography;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.1986.4307780
  • Filename
    4307780