Abstract :
This paper presents a method to decompose multichannel long-term intramuscular electromyogram (EMG) signals. In contrast to existing decomposition methods which only support short registration periods or single-channel recordings of signals of constant muscle effort, the decomposition software EMG-LODEC (ElectroMyoGram LOng-term DEComposition) is especially designed for multichannel long-term recordings of signals of slight muscle movements. A wavelet-based, hierarchical cluster analysis algorithm estimates the number of classes [motor units (MUs)], distinguishes single MUAPs from superpositions, and sets up the shape of the template for each class. Using three channels and a weighted averaging method to track action potential (AP) shape changes improve the analysis. In the last step, nonclassified segments, i.e., segments containing superimposed APs, are decomposed into their units using class-mean signals. Based on experiments on simulated and long-term recorded EMG signals, our software is capable of providing reliable decompositions with satisfying accuracy. EMG-LODEC is suitable for the study of MU discharge patterns and recruitment order in healthy subjects and patients during long-term measurements.
Keywords :
electromyography; medical signal processing; software packages; wavelet transforms; MU discharge patterns; action potential shape changes tracking; healthy subjects; long-term measurements; long-term multichannel EMG signals decomposition; patients; recruitment order; slight muscle movements signals; template shape; wavelet coefficients; wavelet-based hierarchical cluster analysis algorithm; Algorithm design and analysis; Clustering algorithms; Electromyography; Muscles; Recruitment; Shape; Signal design; Software packages; Wavelet analysis; Wavelet coefficients; Action Potentials; Adult; Algorithms; Cluster Analysis; Computer Simulation; Diagnosis, Computer-Assisted; Electromyography; False Negative Reactions; False Positive Reactions; Female; Fingers; Humans; Internet; Male; Middle Aged; Models, Neurological; Monitoring, Ambulatory; Motor Neurons; Movement; Muscle, Skeletal; Musculoskeletal Diseases; Pattern Recognition, Automated; Porphyrins; Reproducibility of Results; Shoulder; Signal Processing, Computer-Assisted; Software;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.2002.807321