• DocumentCode
    1171585
  • Title

    Traceability via the Internet for microwave measurements using vector network analyzers

  • Author

    Dudley, Richard A. ; Ridler, Nick M.

  • Author_Institution
    Centre for Electromagn. & Time Metrol., Nat. Phys. Lab., Teddington, UK
  • Volume
    52
  • Issue
    1
  • fYear
    2003
  • fDate
    2/1/2003 12:00:00 AM
  • Firstpage
    130
  • Lastpage
    134
  • Abstract
    The Internet has been utilized to provide highly efficient and cost-effective measurement services using vector network analyzers (VNAs). Remote calibration and measurement using VNAs from two of the leading instrument manufacturers has been achieved with direct access to primary national measurement standards and procedures, via the Internet, for microwave frequencies in the range 45 MHz to 110 GHz.
  • Keywords
    Internet; calibration; measurement standards; microwave measurement; network analysers; 45 MHz to 110 GHz; Internet; VNAs; microwave measurements; primary national measurement standards; remote calibration; traceability; vector network analyzers; Calibration; IP networks; Instruments; Laboratories; Measurement standards; Metrology; Microwave measurements; Remote monitoring; Transmission line measurements; Web and internet services;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2003.809482
  • Filename
    1191420