  • Abstract
    Silicon drift detectors have been tested for X-ray Spectroscopy applications. By optimizing the detector-FET connection and using a very low leakage current detector manufacturing process, it has been possible to achieve a very good energy resolution. The resolution and leakage current have been studied as a function of temperature, from room temperature down to -30°C, and as a function of active area. Also the effects influencing the peak to background ratio have been outlined
  • Keywords
    X-ray detection and measurement; X-ray spectroscopy; leakage currents; semiconductor counters; 243 to 273 K; Si; Si drift detectors; X-ray spectroscopy; detector-FET connection; energy resolution; low leakage current; peak to background ratio; temperature; Energy resolution; Leak detection; Leakage current; Manufacturing processes; Silicon; Spectroscopy; Temperature; Testing; X-ray detection; X-ray detectors;