DocumentCode
1177098
Title
Development and evaluation of charge-sensitive preamplifier for CsI calorimeter in the KEK B-factory
Author
Tanaka, Manobu ; Ikeda, Hirokazu ; Tamai, Kunio ; Takemoto, Megumi ; Hayashii, Hisashi
Author_Institution
KEK, Nat. Lab. for High Energy Phys., Tsukuba, Japan
Volume
41
Issue
4
fYear
1994
fDate
8/1/1994 12:00:00 AM
Firstpage
1208
Lastpage
1211
Abstract
We developed a charge-sensitive preamplifier for the electromagnetic calorimeter (CsI(Tl)) in the KEK B-factory. We examined noise characteristic, dynamic range and power-supply ripple-rejection ratio on power rails
Keywords
calorimeters; detector circuits; nuclear electronics; preamplifiers; scintillation counters; CsI(Tl) calorimeter; KEK B-factory; charge-sensitive preamplifier; dynamic range; electromagnetic calorimeter; noise characteristics; power rails; power-supply ripple-rejection ratio; Capacitance; Circuits; Electrons; JFETs; Photodiodes; Photonic crystals; Preamplifiers; Semiconductor device noise; Signal to noise ratio; Voltage;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.322885
Filename
322885
Link To Document