Testing 144- and 256-crystal BGO block detectors
Author :
Rogers, Joel G. ; Nutt, Ronald ; Andreaco, Mark ; Williams, C.W.
Author_Institution :
TRIUMF, Vancouver, BC, Canada
fDate :
8/1/1994 12:00:00 AM
Abstract :
New block detectors have been fabricated incorporating large numbers of small crystals. We evaluate and compare the performance of the new detectors with a standard 64-crystal block detector from Siemans-CTI. The new detectors demonstrate greatly improved imaging capability for 511 keV gamma rays. Future PET tomographs incorporating such detectors should produce substantially better volume images with little increase in tomograph manufacturing costs. The detectors will require a new type of automatic calibration procedure. Various such procedures have been tested and are discussed. A technique using higher energy gamma rays has shown special promise
Keywords :
bismuth compounds; calibration; gamma-ray detection and measurement; radioisotope scanning and imaging; scintillation counters; 511 keV; BGO block detectors; Bi4Ge3O12; PET tomographs; automatic calibration; gamma rays; imaging capability; positron emission tomography; tomograph manufacturing costs; Calibration; Costs; Crystals; Gamma ray detection; Gamma ray detectors; Gamma rays; Manufacturing; Optical imaging; Positron emission tomography; Testing;
Journal_Title :
Nuclear Science, IEEE Transactions on