• DocumentCode
    1177508
  • Title

    Testing 144- and 256-crystal BGO block detectors

  • Author

    Rogers, Joel G. ; Nutt, Ronald ; Andreaco, Mark ; Williams, C.W.

  • Author_Institution
    TRIUMF, Vancouver, BC, Canada
  • Volume
    41
  • Issue
    4
  • fYear
    1994
  • fDate
    8/1/1994 12:00:00 AM
  • Firstpage
    1423
  • Lastpage
    1429
  • Abstract
    New block detectors have been fabricated incorporating large numbers of small crystals. We evaluate and compare the performance of the new detectors with a standard 64-crystal block detector from Siemans-CTI. The new detectors demonstrate greatly improved imaging capability for 511 keV gamma rays. Future PET tomographs incorporating such detectors should produce substantially better volume images with little increase in tomograph manufacturing costs. The detectors will require a new type of automatic calibration procedure. Various such procedures have been tested and are discussed. A technique using higher energy gamma rays has shown special promise
  • Keywords
    bismuth compounds; calibration; gamma-ray detection and measurement; radioisotope scanning and imaging; scintillation counters; 511 keV; BGO block detectors; Bi4Ge3O12; PET tomographs; automatic calibration; gamma rays; imaging capability; positron emission tomography; tomograph manufacturing costs; Calibration; Costs; Crystals; Gamma ray detection; Gamma ray detectors; Gamma rays; Manufacturing; Optical imaging; Positron emission tomography; Testing;