DocumentCode :
1180692
Title :
A Versatile Current-Regulated Bidirectional Stimulator
Author :
Caggiano, Vincent J.
Author_Institution :
Biomedical Engineering Group, Division of Neuropsychiatry, Walter Reed Army Institute of Research, Walter Reed Army Medical Center, Washington, D. C.
Issue :
1
fYear :
1965
Firstpage :
32
Lastpage :
39
Abstract :
Based on requirements specified by several investigators engaged in studies involving electrical stimulation through electrodes chronically implanted in brain substance, an electrical stimulator was designed and built. The instrumentation problems encountered in stimulation studies are discussed to emphasize the need for certain features adapting the instrument to the biophysical characteristics of living brain tissues. The resulting stimulator output is bidirectional, current-regulated and isolated, and provides isolated monitors for current ranges from 0.1 mA to 10 mA and pulses from 0.1 ms to 10 ms wide.
Keywords :
Biomedical engineering; Brain injuries; Current control; Electrical stimulation; Electrodes; Instruments; Neurophysiology; Physiology; Production; Propagation delay; Brain; Cats; Electric Stimulation; Engineering;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1965.4502338
Filename :
4502338