• DocumentCode
    1180692
  • Title

    A Versatile Current-Regulated Bidirectional Stimulator

  • Author

    Caggiano, Vincent J.

  • Author_Institution
    Biomedical Engineering Group, Division of Neuropsychiatry, Walter Reed Army Institute of Research, Walter Reed Army Medical Center, Washington, D. C.
  • Issue
    1
  • fYear
    1965
  • Firstpage
    32
  • Lastpage
    39
  • Abstract
    Based on requirements specified by several investigators engaged in studies involving electrical stimulation through electrodes chronically implanted in brain substance, an electrical stimulator was designed and built. The instrumentation problems encountered in stimulation studies are discussed to emphasize the need for certain features adapting the instrument to the biophysical characteristics of living brain tissues. The resulting stimulator output is bidirectional, current-regulated and isolated, and provides isolated monitors for current ranges from 0.1 mA to 10 mA and pulses from 0.1 ms to 10 ms wide.
  • Keywords
    Biomedical engineering; Brain injuries; Current control; Electrical stimulation; Electrodes; Instruments; Neurophysiology; Physiology; Production; Propagation delay; Brain; Cats; Electric Stimulation; Engineering;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1965.4502338
  • Filename
    4502338