DocumentCode
1183202
Title
Calculation and Digital Display of Whole Blood Oxygen Saturation by Analog Techniques
Author
Sutterer, William F. ; Krovetz, L.Jerome
Author_Institution
Waters Company, Rochester, Minn. 55901.
Issue
2
fYear
1969
fDate
4/1/1969 12:00:00 AM
Firstpage
116
Lastpage
122
Abstract
An electronic analog system has been developed which performs the double scale calculation from the Wood oximeter and gives a single output as a log ratio and also gives a digital display of oxygen saturation, with simultaneous recording of these parameters if desired. This system, built of commercially available analog components, consists of operational amplifiers to amplify the output of the red and infrared photocells, logarithmic devices to calculate the optical density as sensed by these two photocells, a division circuit to calculate the log ratio, a scaling circuit to facilitate digital display in percent oxygen saturation, and a digital voltmeter.
Keywords
Blood; Circuits; Digital recording; Displays; Operational amplifiers; Optical amplifiers; Optical recording; Oxygen; Semiconductor optical amplifiers; Stimulated emission; Blood Gas Analysis; Computers; Computers, Analog; Humans; Mathematics; Methods; Oximetry; Oxygen Consumption; Spectrophotometry;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.1969.4502623
Filename
4502623
Link To Document