• DocumentCode
    1184541
  • Title

    An Ultrasound Moving Target Indicator System for Diagnostic Use

  • Author

    Barnes, Ralph W. ; Thurstone, Fredrick L.

  • Author_Institution
    Department of Neurology, Bowman Gray School of Medicine, Wake Forest University, Winston-Salem, N. C. 27103.
  • Issue
    1
  • fYear
    1971
  • Firstpage
    4
  • Lastpage
    8
  • Abstract
    An ultrasound system using moving target indicator (MTI) techniques has been developed to provide a different approach to the problem of isolating and presenting echo information arising from dynamic tissue interfaces. The basic MTI procedure accomplishes isolation of moving target information by comparing echo information subsequent to a transmit pulse to the echo information arising from the preceding pulse. Stationary target information is effectively eliminated. In addition, the MTI procedure isolates dynamic target information throughout the volume investigated by the ultrasonic transducer. This feature eliminates the need for rangegating.
  • Keywords
    Acoustic transducers; Biomedical measurements; Delay effects; Frequency; Heart valves; Pulse amplifiers; Signal processing; System performance; Ultrasonic imaging; Ultrasonic transducers; Biomedical Engineering;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1971.4502782
  • Filename
    4502782