DocumentCode :
1185245
Title :
An Optimally Controlled Respirator
Author :
Mitamura, Yoshinori ; Mikami, Tomohisa ; Sugawara, Hiromi ; Yoshimoto, Chiyoshi
Author_Institution :
Department of Medical Electronics, Research Institute of Applied Electricity, Hokkaido University, Sapporo, Japan.
Issue :
5
fYear :
1971
Firstpage :
330
Lastpage :
338
Abstract :
An optimally controlled respirator was developed. It has three main features: 1) ventilation is controlled by the patient´s metabolic rate from continuously measured C02 output, 2) physiologic dead space approximated as a linear function of tidal volume is used to estimate alveolar ventilation, and 3) respiratory rate is computed to minimize ventilatory work.
Keywords :
Automatic control; Control systems; Dogs; Extraterrestrial measurements; Linear approximation; Medical control systems; Optimal control; Surgery; Ventilation; Volume measurement; Animals; Basal Metabolism; Blood Gas Analysis; Dogs; Humans; Respiration, Artificial; Respiratory Insufficiency; Ventilators, Mechanical;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1971.4502864
Filename :
4502864