DocumentCode :
1185325
Title :
A Self-Centering Respiration Monitor for Small Animal Research
Author :
Graystone, Peter
Author_Institution :
Department of Physiology, University of British Columbia, Vancouver 8, B. C., Canada.
Issue :
5
fYear :
1971
Firstpage :
382
Lastpage :
383
Abstract :
It is necessary to monitor the physiological state of small animals being maintained under anesthetic. One parameter that may be observed is the respiratory pattern which is particularly important when using anesthetics known to depress respiratory function. A respiration monitor is described which indicates nasal air flow and which has a self-centering feature to eliminate drift. This feature enables the sensitivity of the unit to be increased to allow monitoring from animals with very low tidal volumes. The monitor is small and is battery operated to eliminate line frequency interference of sensitive recording equipment. It gives a meter indication of the respiratory pattern and provides an output suitable for driving a recorder. An audible tone is sounded at each inspiration.
Keywords :
Anesthetic drugs; Animals; Batteries; Biomedical monitoring; Capacitors; Feedback circuits; Interference elimination; Operational amplifiers; Resistors; Thermistors; Anesthesia, General; Animals; Monitoring, Physiologic; Rabbits; Respiration;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1971.4502872
Filename :
4502872
Link To Document :
بازگشت