  • Abstract
    We suggest that some key physical parameters of high-purity germanium (HPGe) segmented detectors, including segment capacitance and leakage current, can be derived from standard noise measurements. These parameters are relevant when designing and optimizing the front-end electronics.
  • Keywords
    capacitance measurement; germanium radiation detectors; leakage currents; noise measurement; nuclear electronics; HPGe detector; front-end electronics; high-purity germanium segmented detectors; leakage current; segment capacitance measurement; standard noise measurements; Capacitance; Crosstalk; Detectors; Electrodes; Energy resolution; Germanium; Leak detection; Noise measurement; Preamplifiers; Testing;