Abstract :
We present techniques and systems for a CZT gamma camera head based on an array of discrete CZT detectors and associated front-end electronics. In Section I, we describe the performance of a small 256 detector (16 × 16) CZT imager in terms of homogeneity, energy resolution, and detection efficiency. The array is composed of groups of 16 single detectors (4×4×6 mm3 each), each group served by a 16-channel multiplexed ASIC preamplifier. The electron bi-parametric (BP) spectrum with the depth of interaction information is used to improve detection efficiency. The present work compares the performance (sensitivity, spatial resolution, and contrast) of this small prototype imager and an Anger NaI(Tl) scintillation gamma camera using phantom images with small cold zones in a high-scatter environment. In Section II, we present an approach to the next generation of this technology, a gamma camera with a segmented monolithic CZT detector associated with the electron BP spectrum, which also allows correction for spectral tailing and dramatically improves both detection efficiency and energy resolution. Results with and without BP correction are given for a 10×10×5 mm3 CZT monolithic detector (16 pixels, 2 × 2 mm pixel, and 2.5-mm pitch) associated with 16 individual ASIC preamplifiers: an energy resolution close to 2% FWHM at 122 keV has been obtained on 16 pixels after BP correction.
Keywords :
application specific integrated circuits; biomedical electronics; biomedical imaging; gamma-ray detection; nuclear electronics; position sensitive particle detectors; preamplifiers; semiconductor counters; 10 mm; 16 pixels; 16-channel multiplexed ASIC preamplifier; 2 mm; 4 mm; 5 mm; 6 mm; Anger NaI(Tl) scintillation gamma camera; CZT gamma camera head; CdZnTe-based gamma camera; contrast; detection efficiency; discrete CZT detectors; electron biparametric spectrum; energy resolution; front-end electronics; high-scatter environment; interaction depth; medical imaging; phantom images; prototype imager; segmented monolithic CZT detector; small cold zones; solid-state gamma camera; spatial resolution; spectral tailing; Application specific integrated circuits; Biomedical imaging; Cameras; Electrons; Energy resolution; Gamma ray detection; Gamma ray detectors; Preamplifiers; Sensor arrays; Solid scintillation detectors;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.2004.839070