DocumentCode :
1188782
Title :
Random Sampling Oscillography
Author :
Frye, G.J. ; Nahman, N.S.
Issue :
1
fYear :
1964
fDate :
3/1/1964 12:00:00 AM
Firstpage :
8
Lastpage :
13
Abstract :
A description is given of a sampling oscilloscope technique in which the sampling pulse recurrence rate is not in any way synchronized with the signal waveform. The technique eliminates the need for a signal channel delay line. An experimental system which employs only semiconductor devices is described, and experimental data demonstrating the practicality of the system is presented.
Keywords :
Bandwidth; Delay lines; Instruments; Klystrons; Laboratories; Oscilloscopes; Pulse amplifiers; Pulse circuits; Pulse generation; Sampling methods;
fLanguage :
English