DocumentCode
1190213
Title
A Precision Dc Potentiometer Microwave Insertion Loss Test Set
Author
Stelzried, C.T. ; Reid, M.S. ; Petty, S.M.
Volume
15
Issue
3
fYear
1966
Firstpage
98
Lastpage
104
Abstract
A simple, accurate microwave test set has been devised for measuring coaxial or waveguide insertion loss. The instrument is designed primarily for highest precision with low attenuations. It is composed almost entirely of commercially available equipment and components, is portable, and can be operated in the laboratory or in the field from ac mains or batteries. The precision of the measurements depends on the connector type and value of the component measured. The precision is better than 0.0001 dB when measuring low insertion losses, discounting the error due to the connectors. Equipment nonlinearity reduces the precision in proportion to the measured attenuation in dB. The frequency range covered is 10 MHz to 40 GHz, which can be extended to 90 GHz with slight perfornance degradation.
Keywords
Attenuation; Battery charge measurement; Coaxial components; Connectors; Insertion loss; Loss measurement; Microwave devices; Microwave measurements; Potentiometers; Testing;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1966.4313517
Filename
4313517
Link To Document