DocumentCode
1194957
Title
Power deposition inside a phantom for testing of MRI heating
Author
Amjad, A. ; Kamondetdacha, R. ; Kildishev, A.V. ; Park, S.M. ; Nyenhuis, J.A.
Author_Institution
Sch. of Electr. & Comput. Eng., Purdue Univ., West Lafayette, IN, USA
Volume
41
Issue
10
fYear
2005
Firstpage
4185
Lastpage
4187
Abstract
A patient in MRI is exposed to the time varying gradient and radio-frequency (RF) fields. Measurements of RF-induced temperature rise in implant in phantoms are used to predict the in-vivo temperature rise. This paper focuses on the numerical computation of power deposition inside such a phantom. The finite-difference time domain (FDTD) was used for the calculations. The phantom was placed inside a whole body bird cage coil. For landmarks in the torso of the phantom, the power deposition is concentrated near the edges. The whole phantom average specific absorption rate (SAR) for a mean square field intensity of 1(μT)2 ranged from 0.032 W/kg for the eyes landmark to 0.151 W/kg for the waist landmark. The SAR increased with increasing conductivity in the range 0.2 to 0.8 S/m. The distribution of the measured temperature rise versus lateral position at the landmark was consistent with the calculation. The electric field is tangential to the phantom edges but has significant elliptical polarization near the corners.
Keywords
biomedical MRI; dosimeters; finite difference time-domain analysis; phantoms; temperature measurement; MRI heating testing; RF magnetic field; RF-induced temperature rise; bird cage coil; finite-difference time domain method; phantoms; power deposition; specific absorption rate; Birds; Finite difference methods; Heating; Imaging phantoms; Implants; Magnetic resonance imaging; Radio frequency; Temperature measurement; Testing; Time domain analysis; FDTD; MRI; RF magnetic field; heating;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/TMAG.2005.854840
Filename
1519575
Link To Document