DocumentCode
1195394
Title
Automatic Hold-Off Adjustment to Avoid False Triggering of Oscilloscope Time Bases
Author
Bertolaccini, M. ; Cova, S.
Volume
22
Issue
1
fYear
1973
fDate
3/1/1973 12:00:00 AM
Firstpage
92
Lastpage
94
Abstract
A system is described that avoids false synchronization of the trigger on signals arriving while the hold-off circuit is resetting, thus preventing the apparent shifting of displayed waveforms. The effectiveness of the system is illustrated by results obtained with different oscilloscopes and signals.
Keywords
Circuit stability; Delay effects; Frequency synchronization; Inspection; Instruments; Manufacturing automation; Nuclear power generation; Oscilloscopes; Switches; Trigger circuits;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1973.4314105
Filename
4314105
Link To Document