DocumentCode
1195851
Title
The Reduction of Errors in a Precise Microwave Attenuator Calibration System
Author
Hollway, D.L. ; Somlo, P.I.
Volume
22
Issue
3
fYear
1973
Firstpage
268
Lastpage
270
Abstract
Refinements have been made to the parallel-substitution heterodyne method of attenuation measurement. These are provision for small fixed frequency differences between the IF signals to improve the sensitivity of tests for leakage, a switched reference path to reduce errors from drift and to assess "flange scatter," a flexible waveguide in the measuring path to accommodate attenuators of different lengths, a steady flow of dry air through the system to remove moisture, and an electronic system for fine-balancing the two channels.
Keywords
Attenuation measurement; Attenuators; Calibration; Electronic equipment testing; Flanges; Fluid flow measurement; Frequency measurement; Length measurement; Scattering; System testing;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1973.4314161
Filename
4314161
Link To Document