DocumentCode
1195953
Title
Time-Domain Reflectometer Measurement of Insertion Loss of High-Frequency Switches
Author
Cormack, George D.
Volume
22
Issue
4
fYear
1973
Firstpage
291
Lastpage
295
Abstract
High-frequency switches of the reed relay type have insertion losses of the order of 0.1 dB. The insertion loss is primarily the result of large skin effect impedance in the magnetically active portion of the signal path. This paper presents a novel time-domain reflectometer technique for measuring this skin effect impedance in the time domain then shows how a Fourier transform will convert the readings to either an impedance or insertion loss in the frequency domain. The method uses inexpensive equipment, is very accurate, and is fast since only a single voltage measurement at one instant of time is needed for the determination of the insertion loss at a desired frequency. Precision of at least 0.001 dB is achievable.
Keywords
Frequency measurement; Impedance measurement; Insertion loss; Loss measurement; Magnetic domains; Relays; Skin effect; Switches; Time domain analysis; Time measurement;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1973.4314175