• DocumentCode
    1196786
  • Title

    International Comparison of Complex Permittivity Measurement at 9 GHz

  • Author

    Bussey, Howard E. ; Morris, Derek ; Zal´tsman, E.B.

  • Volume
    23
  • Issue
    3
  • fYear
    1974
  • Firstpage
    235
  • Lastpage
    239
  • Abstract
    Dielectric constant and loss measurements made by three government laboratories (of the USSR, Canada, and the U. S. A.) are compared. The two materials measured were glasses. The measurements utilized cavity resonators in the H01n mode. The errors in dielectric constant reported by the laboratories were usually ± 0.3 percent; actual differences between laboratories of the average corrected results were only ± 0.05 percent. The loss tangent results disagreed when multimoding occurred; however, the errors may be as low as ± 0.00002 or ± 3 percent, whichever is the larger, if multimoding is avoided.
  • Keywords
    Cavity resonators; Dielectric constant; Dielectric loss measurement; Dielectric materials; Dielectric measurements; Glass; Government; Laboratories; Loss measurement; Permittivity measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1974.4314270
  • Filename
    4314270