DocumentCode
1197009
Title
Standard Cell Calibration via Current Transfer
Author
Williams, Edwin R. ; Olsen, Paul T. ; Field, Bruce F.
Volume
23
Issue
4
fYear
1974
Firstpage
299
Lastpage
302
Abstract
The EMF´s of standard cells are now being transferred between laboratories over a 1¿-km cable with a precision of 4 parts in 108 to provide an instantaneous comparison of the 2e/h and ¿p¿ experiments being carried out at the two facilities. This is accomplished by transferring a constant current that produces a 1-V drop across standard resistors located at both ends of the cable.
Keywords
Calibration; Current measurement; Laboratories; Magnetic field measurement; Monitoring; NIST; Nuclear magnetic resonance; Protons; Resistors; Voltage;