• DocumentCode
    1198452
  • Title

    A Wide-Range High-Voltage Capacitance Bridge with One PPM Accuracy

  • Author

    Petersons, Oskars ; Anderson, William E.

  • Volume
    24
  • Issue
    4
  • fYear
    1975
  • Firstpage
    336
  • Lastpage
    344
  • Abstract
    An impedance bridge for high-voltage capacitance and related measurements to an accuracy of 1 ppm is described and the background research and development which made it possible is documented. The bridge is of the transformer-ratioarm type, the principal components of which are a comparator and several two-stage transformers. The bridge can be used to measure capacitance ratios over a range from 1/1 to 107/1 with a resolution between 0.1 and 0.25 ppm. The highest accuracies are obtained at the principal power frequencies of 50-60 Hz, but the bridge is usable up to 400 Hz. The factors which limited the accuracy of previously developed bridges of this type were reexamined and their influence reduced. Two independent methods were developed for the calibration of the bridge.
  • Keywords
    Bridge circuits; Calibration; Capacitance measurement; Capacitors; Impedance measurement; Instruments; Laboratories; NIST; Research and development; Transformers;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1975.4314455
  • Filename
    4314455