• DocumentCode
    1200025
  • Title

    Improvements in a Calorimeter for High-Power CW Lasers

  • Author

    Chamberlain, George E. ; Simpson, Philip A. ; Smith, Richard L.

  • Volume
    27
  • Issue
    1
  • fYear
    1978
  • fDate
    3/1/1978 12:00:00 AM
  • Firstpage
    81
  • Lastpage
    86
  • Abstract
    Measurement certainty with the BB series of electrically calibrated calorimeters for high-energy lasers has been enhanced by the addition of monitors for energy backscattered from the meter and for energy missing the entrance aperture (overspill). The performance and design features of the recently constructed BB2 meter are compared with the previously described BB1 meter. Direct intercomparison shows the agreement between meters to be 1 percent.
  • Keywords
    Apertures; Backscatter; Calibration; Energy measurement; Laser beams; Laser theory; NIST; Optical design; Optical receivers; Optical sensors;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1978.4314625
  • Filename
    4314625