DocumentCode :
1200232
Title :
Programmable Automatic Multimeter Calibration System
Author :
Warner, Eliot M. ; West, James L.
Volume :
27
Issue :
2
fYear :
1978
fDate :
6/1/1978 12:00:00 AM
Firstpage :
156
Lastpage :
159
Abstract :
This paper describes a programmable automatic ac and dc multimeter calibration system consisting of a remotely programmed precise source of ac and dc voltage and current as well as resistance and a desk calculator. The calculator is used not only to sequence the source through different test conditions but also to display information, determine conformity to limits, and record data. Emphasis is placed on reducing the difficulty of use by an operator both in the calibration function and in the data entry function. Interactive routines for accomplishing these functions are described.
Keywords :
Automatic control; Calibration; Displays; Frequency; Humans; Immune system; Inspection; Instruments; System testing; Voltage;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1978.4314648
Filename :
4314648
Link To Document :
بازگشت