DocumentCode :
1201616
Title :
A Universal Four-Pair Impedance Bridge
Author :
Corney, Andrew C.
Volume :
28
Issue :
3
fYear :
1979
Firstpage :
211
Lastpage :
215
Abstract :
A wide range 10 ppm four-terminal-pair audiofrequency impedance bridge is described. The bridge can be operated either in an impedance mode, for self-inductance, mutual inductance, and resistanre measurements; or in an admittance mode for capacitance and conductance measurements. The frequency range for full aecuracy is 40 Hz to 2 kHz. Maintenance of the calibration of the bridge is facilitated by the use of transformer ratios to extend the range of a small number of precision components, and extensive use is made of electronic techniques to automate auxiliary balances.
Keywords :
Admittance measurement; Bridge circuits; Calibration; Capacitance measurement; Cryogenics; Electrical resistance measurement; Frequency; Impedance measurement; Inductance measurement; Physics;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1979.4314807
Filename :
4314807