• DocumentCode
    1201828
  • Title

    The design and construction of high field-uniformity permanent magnet system for MRI

  • Author

    Feng, Z.X. ; Jiang, X.H. ; Han, S.

  • Author_Institution
    Inst. of Electr. Eng., Acad. Sinica, Beijing, China
  • Volume
    28
  • Issue
    1
  • fYear
    1992
  • fDate
    1/1/1992 12:00:00 AM
  • Firstpage
    641
  • Lastpage
    643
  • Abstract
    It is shown that ferromagnetic pole pieces can effectively improve the field uniformity of an MRI (magnetic resonance imaging) device. First, the specially designed ferromagnetic pole pieces are placed onto the surfaces of the main permanent-magnet poles to improve the field uniformity preliminarily. Then, the necessary field uniformity is obtained by using the magneto-dipoles which are suitably placed in the field domain. The specially designed pole pieces can increase the field-uniformity by two orders of magnitude
  • Keywords
    biomedical NMR; biomedical equipment; magnetic resonance spectrometers; permanent magnets; spectrometer components and accessories; MRI; construction; design; ferromagnetic pole pieces; high field-uniformity; magneto-dipoles; permanent magnet system; Coils; Costs; Magnetic devices; Magnetic domains; Magnetic fields; Magnetic materials; Magnetic resonance imaging; Permanent magnets; Saturation magnetization; Superconducting magnets;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.119959
  • Filename
    119959