DocumentCode
1202663
Title
Accurate Absolute Frequency Measurements on Stabilized CO2 and He-Ne Infrared Lasers
Author
Clairon, Andre ; Dahmani, Brahim ; Rutman, Jacques
Volume
29
Issue
4
fYear
1980
Firstpage
268
Lastpage
272
Abstract
In our laboratory, we have measured the frequencies of CO2 and He-Ne lasers near 30 and 88 THz, stabilized, respectively, by saturated fluorescence in CO2 and saturated absorption in CH4. Our measurement system includes a stable free-running optically pumped CH3OH laser at 4.25 THz replacing the noisy H2O laser used as a transfer oscillator in early experiments. As a result of the reduced mixing orders (¿9), beat notes between lasers are now observed with ¿30-dB signal-to-noise (S/N) ratios in a 100-kHz bandwidth. Therefore, beat frequencies can be measured accurately with digital counters and simultaneous counting of the frequencies involved largely eliminates the uncertainties due to transfer oscillators. The measurements are referred to the cesium beam frequency standard. The results are processed by a desktop calculator which also controls the measurement process.
Keywords
Electromagnetic wave absorption; Fluorescence; Frequency measurement; Laboratories; Laser stability; Optical mixing; Optical pumping; Optical saturation; Oscillators; Pump lasers;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1980.4314932
Filename
4314932
Link To Document