  • Abstract
    Methods are described for measuring the settling times and other dynamic characteristics of voltage and current output D/A converters (DAC´s), operational amplifiers, and precision voltage step generators. Circuits are described for measuring voltage-output device settling times as short as 1¿s to within a settling error of ±2 ppm, and current-output device settling times as short as 40 ns to within a settling error of ±0.012 percent.
  • Keywords
    Bandwidth; Calibration; Character generation; Circuits; Current measurement; Operational amplifiers; Oscilloscopes; Sampling methods; Time measurement; Voltage;