Title :
On the Measurement Accuracy of the Tissue Permittivity in Vivo
Author :
Kraszewski, Andrzej ; Stuchly, Stanislaw S. ; Stuchly, Maria A. ; Symons, Stephen A.
fDate :
3/1/1983 12:00:00 AM
Abstract :
An open-ended coaxial line and a computer-controlled network analyzer have been extensively used for measuring the tissue permittivity in vivo at radio and microwave frequencies. An analysis is presented of measurement errors resulting from calibration and random errors of the network analyzer [1],[2]. The uncertainties in our system are ¿¿ = 0.05 dB and ¿¿ = 0.3°. Key Words-Radio frequency, Microwave permittivity in vivo, Automated network analyzer.
Keywords :
Calibration; Coaxial components; Computer errors; Computer networks; Frequency measurement; In vivo; Measurement errors; Microwave frequencies; Microwave measurements; Permittivity measurement;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.1983.4315006