• DocumentCode
    1203394
  • Title

    Comparison between the 127I2 Stabilized He-Ne Lasers at 633 nm and 612 nm of the BIPM and the IMGC

  • Author

    Bertinetto, F. ; Cordiale, P. ; Picotto, G.B. ; Chartier, Jean-Marie ; Felder, R. ; Glaser, M.G.

  • Volume
    32
  • Issue
    1
  • fYear
    1983
  • fDate
    3/1/1983 12:00:00 AM
  • Firstpage
    72
  • Lastpage
    76
  • Abstract
    In November 1981, a comparison between BIPM and IMGC He-Ne lasers stabilized to saturated absorption lines of 127I2, at the wavelengths ¿ = 633 nm and ¿ = 612 nm has been made at the BIPM. At 633 nm, a repeatability and a reproducibility within +1 × 10-11 · ¿ have been obtained. At 612 nm, a repeatability within a few 10-11 · ¿ only for one laser and a reproducibility of ± 2 × 10-10 · ¿ have been obtained.
  • Keywords
    Absorption; Argon; Curve fitting; Frequency; Laboratories; Laser stability; Least squares methods; Power lasers; Reproducibility of results; Testing;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1983.4315012
  • Filename