• DocumentCode
    1203807
  • Title

    The Maintenance and Comparison of Standards of Electromotive Force at the BIPM

  • Author

    Witt, Thomas J. ; Reymann, Dominique

  • Volume
    32
  • Issue
    1
  • fYear
    1983
  • fDate
    3/1/1983 12:00:00 AM
  • Firstpage
    260
  • Lastpage
    266
  • Abstract
    This article describes improved standard cell comparison equipment, temperature-controlled enclosures, and measurement techniques used to relate reference groups of high-quality standard cells to the BIPM (Bureau International des Poids et Mesures) Josephson-effect standard with an uncertainty of less than 10 nV. This performance is demonstrated by results obtained over the last four years. The same techniques are now being applied in the routine comparison of international standards of electromotive force by the transfer of standard cells. An analysis of the behavior of the cells that have participated in comparisons over the last 12 years indicates that the uncertainty due to the instability of cells as transfer standards is about 0.5 ¿V, i.e., one hundred times greater than the best reproducibility in the laboratory. A comprehensive tabulation of uncertainties shows this to be by far the greatest single contribution.
  • Keywords
    Calibration; Force measurement; Frequency conversion; Histograms; Josephson effect; Laboratories; Measurement standards; Measurement techniques; Reproducibility of results; Uncertainty;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1983.4315055
  • Filename
    4315055