• DocumentCode
    1203964
  • Title

    Coherent Optical Recognition and Counting of Reticulated Red Blood Cells

  • Author

    Ward, Joseph E. ; Carlson, F.Paul ; Heywood, J.David

  • Author_Institution
    Central Research Laboratories, 3M Company, St. Paul, Minn. 55133.
  • Issue
    1
  • fYear
    1974
  • Firstpage
    12
  • Lastpage
    20
  • Abstract
    A coherent optical data-processing system for identifying and counting blood cells has been developed. Optical Weiner-Kolmogorov (minimum mean-square error) linear filtering is employed to estimate the density per unit area of a given cell type present in a sample blood smear. The Weiner filter is designed to recognize the characteristics of an ensemble of cells of a type rather than specific individual cells. In this way the system is made insensitive to minor variations in shape, size, or orientation of the cells. A Poisson shot-noise model for blood cells as a spatial random process is employed to calculate the far-field scattering from the cells. Results of experiments in which the percentage of reticulated (immature) red cells present in a smear has been determined are presented. The identification and counting accuracies compare favorably with those obtained by laboratory technicians with a microscope. The potential time saving of instantaneous estimation for the entire smear is significant.
  • Keywords
    Cells (biology); Character recognition; Laboratories; Maximum likelihood detection; Nonlinear filters; Optical filters; Optical scattering; Random processes; Red blood cells; Shape; Automatic Data Processing; Erythrocyte Count; Filtration; Mathematics; Methods; Models, Biological; Optics;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1974.324356
  • Filename
    4120712