• DocumentCode
    1204224
  • Title

    Development of a Korotkov Sound Processor for Automatic Identification of Auscultatory Events - Part I: Specification of Preprocessing Bandpass Filters

  • Author

    Golden, Donald P. ; Wolthuis, Roger A. ; Hoffler, G.Wyckliffe ; Gowen, Richard J.

  • Author_Institution
    Department of Management, Health and Hospitals Governing Commission of Cook County, Chicago, Ill. 60612.
  • Issue
    2
  • fYear
    1974
  • fDate
    3/1/1974 12:00:00 AM
  • Firstpage
    114
  • Lastpage
    118
  • Abstract
    Frequency bands that best discriminate the Korotkov sounds at systole and at diastole from the sounds immediately preceding these events are defined. Korotkov sound data were recorded from five normotensive subjects during orthostatic stress [lower body. negative pressure (LBNP)] and bicycle ergometry. A spectral apalysis of the seven Korotkov sounds centered about the systolic and diastolic auscultatory events revealed that a maximum increase in amplitude at the systolic transition occurred in the 18-26-Hz band, while a maximum decrease in amplitude at the diastolic transition occurred in the 40-60-Hz band. These findings were remarkably consistent across subjects and test conditions. These passbands are included in the design specifications for an automatic blood pressure. measuring system used in conjunction with medical experiments during NASA´s Skylab program.
  • Keywords
    Band pass filters; Biomedical measurements; Blood pressure; Frequency; Laboratories; Logic circuits; NASA; Passband; Pressure measurement; Space technology; Auscultation; Automation; Biomedical Engineering; Blood Pressure Determination; Humans;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1974.324296
  • Filename
    4120741