Title :
Comparison of quantized hall resistance with a 1-Ω standard
Abstract :
A major step in the precise determination of the quantized Hall resistance is the comparison of a resistor of value nominally 6453.2 Ω with a 1-Ω standard. For this purpose the National Measurement Laboratory will use a build-up resistor comprising 83 resistors each of 801/3 Ω, giving a ratio of 64534/9. An accuracy of better than 1 in 108 is expected.
Keywords :
Calibration; Fluctuations; Frequency measurement; Noise measurement; Noise reduction; Oscillators; Phase measurement; Resistors; Stability; Testing;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.1983.4315104