DocumentCode :
1204271
Title :
Localized DC Field Produced by Diode Implanted in Isotropic Homogeneous Medium and Exposed to Uniform RF Field
Author :
Schuder, John C. ; Gold, Jerry H.
Author_Institution :
Thoracic and Cardiovascular Section, Department of Surgery, University of Missouri, Columbia, Mo. 65201.
Issue :
2
fYear :
1974
fDate :
3/1/1974 12:00:00 AM
Firstpage :
152
Lastpage :
163
Abstract :
Minimally invasive selective stimulation of biological tissue within the body can be achieved by implanting a small rectifying diode and then applying an RF field by means of electrodes or a coil on or near the surface of the body. A theoretical analysis of a simplified model consisting of a cylindrically symmetrical unit implanted in an isotropic homogeneous medium of conductivity ¿and permittivity ¿ and exposed to a uniform RF field relates the dc component of electrode current to the detailed geometry of the electrodes of the implanted unit and to the amplitude of the applied field.
Keywords :
Biological system modeling; Biological tissues; Coils; Conductivity; Diodes; Electrodes; Minimally invasive surgery; Permittivity; Radio frequency; Solid modeling; Electric Stimulation; Electrodes, Implanted; Electronics, Medical; Humans; Mathematics; Models, Biological;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1974.324301
Filename :
4120746
Link To Document :
بازگشت