DocumentCode :
1204484
Title :
Eight-Channel Data Set for Clinical EEG Transmission Over Dial-Up Telephone Network
Author :
Gardner, Reed M. ; Bennet, Donald R. ; Vorce, Richard B.
Author_Institution :
Department of Biophysics and Bioengineering, Department of Neurology, University of Utah, Salt Lake City. Utah 84103.
Issue :
3
fYear :
1974
fDate :
5/1/1974 12:00:00 AM
Firstpage :
246
Lastpage :
249
Abstract :
Telephone telemetry is becoming increasingly more applicable to health care delivery. Technical characteristics and results of field testing of a new eight-channel constant-bandwidth (CBW) frequency-modulated data set are described in this communication. These tests have shown that eight channels of quality EEG can now be successfully transmitted simultaneously over one conventional telephone line. Economic considerations and other applications are discussed.
Keywords :
Calibration; Circuit testing; Electroencephalography; FCC; Frequency; Insulation; Oscilloscopes; Pulse shaping methods; Telemetry; Telephony; Electroencephalography; Technology, Medical; Telemetry; Telephone;