Abstract :
A percutaneous wire electrode has been developed which permits long-term recording of intramuscular EMG signals. The electrode consists of a small coil wound from 0.0011 inch diameter wire and filled with silicone rubber. Easily inserted by means of a hypodermic needle, it has demonstrated reliable EMG recording for periods of up to thirteen months.
Keywords :
Acceleration; Electrodes; Electromyography; Equations; Legged locomotion; Muscles; Needles; Physiology; Skin; Wire; Animals; Dogs; Electrodes, Implanted; Electromyography; Humans; Skin;