Title :
The SVEC III vectorcardiographic lead system
Author :
Malmivuo, Jaakko A.
Author_Institution :
Ragnar Granit Inst., Tampere, Finland
Abstract :
The properties of the stereovectorelectrocardiography (SVEC) III vectorcardiographic (VCG) lead system designed by Schmitt and Simonson are described. The theoretical and clinical aspects of this system are compared with the two other widely applied systems, of Frank and McFee. The theory and short history of VCG recording are also discussed.
Keywords :
bioelectric potentials; electrocardiography; stereovectorelectrocardiography III vectorcardiographic lead system; Bioelectric phenomena; Disk recording; Displays; Electrocardiography; Electrodes; Heart; History; Mirrors; Reproducibility of results; Thorax;
Journal_Title :
Engineering in Medicine and Biology Magazine, IEEE
DOI :