DocumentCode :
1206568
Title :
The Importance of an Air Pressure Seal on Standard Cell Enclosures with High Temperature Stability
Author :
Clark, Richard F. ; Jurkus, Algirdas P. ; Mclaren, George D. ; Wood, Barry M.
Issue :
3
fYear :
1985
Firstpage :
468
Lastpage :
470
Abstract :
The use of a pressure seal on standard cell still-air enclosures having high-temperature stability is discussed. The seal eliminates the effect of air pressdre variations on the temperature of the cells and the temperature monitor. It is calculated that a pressure change of 1 percent for air at normal temperature and pressure will change the air temperature 0.85 K under adiabatic conditions. While the conditions in a still-air enclosure are far from adiabatic, a measurement made on an oven under development gave an air inner chamber temperature change of 0.8 mK for this pressure change.
Keywords :
Circuit testing; Counting circuits; Instruments; Linearity; Microprocessors; Performance evaluation; Seals; Stability; Temperature measurement; Temperature sensors;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1985.4315373
Filename :
4315373
Link To Document :
بازگشت