DocumentCode
1206653
Title
Optical Constants of a Magnetic-Type Electron Microscope
Author
Marton, Lorinc ; Hutter, R.G.E.
Volume
32
Issue
9
fYear
1944
Firstpage
546
Lastpage
552
Abstract
The majority of transmission-type electron microscopes use magnetic electron lenses because of their superiority with respect to resolving power as compared with that of electrostatic lenses. The axial field distribution of magnetic lenses customarily in use can be approximated very closely by the analytical expression H(z) = H0 /[I+(z/a)2]. H(z) is the field strength along the optical axis z. H0 is the maximum field strength at the point z = 0. 2a is the so-called half width of the field curve, i.e., H(±a) = H0 /2. The paraxial-ray differential equation can be rigorously solved for such a field. On the basis of this solution, it is possible to calculate accurately all optical constants which are of interest in the design and performance of a microscope. Previous measurements and calculations are summarized and extended to give a complete description of the behavior of this type of electron microscope.
Keywords
Differential equations; Electron microscopy; Electron optics; Electrostatics; Lenses; Magnetic analysis; Magnetic force microscopy; Optical design; Optical microscopy; Transmission electron microscopy;
fLanguage
English
Journal_Title
Proceedings of the IRE
Publisher
ieee
ISSN
0096-8390
Type
jour
DOI
10.1109/JRPROC.1944.233069
Filename
1695067