• DocumentCode
    1206702
  • Title

    An Evaluation of Josephson-Effect Voltage Standard for the US Army: A Preliminary Report

  • Author

    Becker, Latika S.R.

  • Issue
    4
  • fYear
    1985
  • Firstpage
    504
  • Lastpage
    506
  • Abstract
    The experimental procedure used and the results obtained on the evaluation of the Josephson-Effect Voltage (JEV) Standard at the US Army Primary Standards Laboratory (APSL) are reported. The inherent precision of measurement is 0.62 ppm (3??). The accuracy of the final result which is determined by various uncertainties is about 1 ppm. The Josephson junction (J-J) is used as a frequency-to-voltage converter with a conversion factor equal to 2e/h. The adopted value of 2e/h is 483593.420 GHz/V(NBS).
  • Keywords
    Josephson junctions; Laboratories; Law; Legal factors; Measurement standards; Microstrip; NIST; Potentiometers; Switches; Voltage measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1985.4315391
  • Filename