• DocumentCode
    1207051
  • Title

    The Characterization of Transcutaneous Stimulating Electrodes

  • Author

    Brennen, Kenneth R.

  • Author_Institution
    Neurological Division, Medtronic, Inc., Minneapolis, MN 55418.
  • Issue
    4
  • fYear
    1976
  • fDate
    7/1/1976 12:00:00 AM
  • Firstpage
    337
  • Lastpage
    340
  • Abstract
    A simple mathematical model of a transcutaneous stimulating electrode molded from conductive elastomer is used to predict the current density pattern produced by such electrodes. A general method has been developed to determine experimentally the current distribution produced by this class of electrodes. Preliminary experimental results are qualitatively in agreement with predictions made from the model. The experimental method for the determination of current spread yields a valid assessment of electrode performance independent of the accuracy of the mathematical model.
  • Keywords
    Biomedical electrodes; Current distribution; Equations; Immune system; Impedance; Mathematical model; Pain; Rubber; Skin; Strips; Electrodes; Electrophysiology; Humans; Mathematics; Models, Biological; Skin;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1976.324594
  • Filename
    4121059