• DocumentCode
    1207290
  • Title

    Extended Application of a Cobalt Glass Dosimeter

  • Author

    Johnson, William T.K.

  • Author_Institution
    Army Materiel Command Harry Diamond Laboratories Washington, D. C.
  • Volume
    10
  • Issue
    4
  • fYear
    1963
  • Firstpage
    23
  • Lastpage
    29
  • Abstract
    Cobalt glass dosimeters, normally used to measure gamma ray and electron fields, have been used to measure gamma ray and coexistent thermal neutron fields in the presence of a fast neutron field. The dosimeter is darkened in proportion to the energy deposited and this darkening is calibrated in known gamma ray and thermal neutron fields. A three-glass sandwich was used to distinguish the effects of gamma rays and thermal neutrons. The sensitivity of the glass to various energy gamma and x-rays is presented. A simple spectrophotometer is used to measure the transmission of light through the glass dosimeter.
  • Keywords
    Calibration; Cobalt; Dosimetry; Electrons; Gamma ray effects; Glass; Laboratories; Neutrons; Stability; Wavelength measurement;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS2.1963.4315459
  • Filename