• DocumentCode
    1207532
  • Title

    Cobalt Glass Dosimetry

  • Author

    Friddell, K.D.

  • Author_Institution
    Boeing Radiation Effects Laboratory the Boeing Company
  • Volume
    11
  • Issue
    5
  • fYear
    1964
  • Firstpage
    155
  • Lastpage
    163
  • Abstract
    The fading of radiation-induced absorbancy of cobalt glass as a function of exposure rate and time was investigated. A linear accelerator (Linac) and a 600 curie cobalt-60 source were used. The fading of the absorbancy was found to obey an empirical relationship over four orders of magnitude of the form A(t) = A(to) - n In t/to where A(to) is the absorbancy at time to. The slopes of the fading curves (n) were found to depend upon the initial absorbancy A(to) and wavelength. The difficulties of using a calibration curve determined with a low-exposure rate source for Linac dosimetry are briefly discussed.
  • Keywords
    Calibration; Cobalt; Dosimetry; Fading; Glass; Ionizing radiation; Linear accelerators; Linear particle accelerator; Semiconductor device measurement; Wavelength measurement;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS2.1964.4315486
  • Filename