Title :
A New High-Gain Multiplier Phototube for Scintillation Counting
Author :
Widmaier, W. ; Engstrom, R.W. ; Stoudenheimer, R.G.
Author_Institution :
Radio Corp. of America, Lancaster, Pa.
Abstract :
This paper describes the RCA-6810, a head-on type of multiplier phototube having an S-11 response and intended for use in high-speed scintillation counting. This tube features a 2-inch-diameter, semitransparent cathode, and has a fourteen-stage, in-line, electrostatically-focused, silver-magnesium dynode structure. The mechanical construction of the tube and the method of assembly are described. With a supply voltage of 2000 volts, the 6810 is capable of multiplying feeble pulses 12.5 million times and has a luminous sensitivity of 750 amperes per lumen. With the maximum rated supply voltage of 2300 v, the amplification factor is 66 million and the luminous sensitivity 4000 a per lumen. Output current pulses of 0.7 a may be utilized. Data showing the linearity of the output over a wide range of light levels are given. Other properties of the tube such as speed of response, after-pulses, and time resolution capability are discussed, and the effect of a magnetic field on performance is described.
Keywords :
Beta rays; Chemicals; Crystals; Electrons; Gamma rays; Laboratories; Photoelectricity; Pulse amplifiers; Radiation detectors; Voltage;
Journal_Title :
Nuclear Science, IRE Transactions on
DOI :
10.1109/TNS2.1956.4315565