Title
Dual Microcomputer Analysis of Cardiac Transmembrane Action Potentials
Author
Klein, Debbie-Lynn ; Jenkins, Janice M. ; Ten Eick, Robert E.
Author_Institution
G. D. Searle and Co.
Issue
12
fYear
1983
Firstpage
819
Lastpage
825
Abstract
A dual microcomputer system has been developed for automating the analysis of cardiac transmembrane action potentials. This system consists of two microcomputers and supporting function modules which digitize, detect, analyze, and plot the transmembrane action potentials from a cardiac cell impaled with a microelectrode. The action potentials are digitized at two sampling rates for computer processing; a sampling rate of 10 kHz is used to obtain the rapid initial upstroke, while a rate of 100 Hz is used to obtain the slower repolarization phase. The action potentials are characterized by graphic displays using the derived measurements. The system measures the resting potential, overshoot amplitude, dV/dtmax and time for 50, 70, and 95 percent repolarization of the action potential.
Keywords
Analog computers; Biology computing; Computer displays; Data acquisition; Hardware; Magnetic materials; Microcomputers; Microelectrodes; Microprocessors; Sampling methods; Action Potentials; Biomedical Engineering; Cell Membrane; Computers; Heart; Microcomputers;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on