Title
An Ultrafiltration Monitor for Hemodialysis Research
Author
Gisser, David ; Strait, Gary ; Zelman, Allen ; Bastidas, Victor ; Harrow, Jeffrey ; Kablitz, Carl ; Stephen, Robert
Author_Institution
Department of Electrical Computer, and Systems Engineering, Rensselaer Polytechnic Institute
Issue
2
fYear
1983
Firstpage
132
Lastpage
136
Abstract
The ultrafitration rate of hemodialysis membranes should be carefully controlled for maximum patient benefit. A special purpose instrument is described that takes in analog pressure information and the analog output of an appropriate weighing system and produces both digital displays and analog outputs (for a strip-chart recorder) of ultrafiltration rate and ultrafiltration index as well as the averaged mean transmembrane hydrostatic pressure drop and volume of ultrafiltrate. The instrument normally uses an averaging time of about 2 min.