DocumentCode :
1211408
Title :
An Ultrafiltration Monitor for Hemodialysis Research
Author :
Gisser, David ; Strait, Gary ; Zelman, Allen ; Bastidas, Victor ; Harrow, Jeffrey ; Kablitz, Carl ; Stephen, Robert
Author_Institution :
Department of Electrical Computer, and Systems Engineering, Rensselaer Polytechnic Institute
Issue :
2
fYear :
1983
Firstpage :
132
Lastpage :
136
Abstract :
The ultrafitration rate of hemodialysis membranes should be carefully controlled for maximum patient benefit. A special purpose instrument is described that takes in analog pressure information and the analog output of an appropriate weighing system and produces both digital displays and analog outputs (for a strip-chart recorder) of ultrafiltration rate and ultrafiltration index as well as the averaged mean transmembrane hydrostatic pressure drop and volume of ultrafiltrate. The instrument normally uses an averaging time of about 2 min.
Keywords :
Artificial biological organs; Biomedical engineering; Biomembranes; Blood; Cities and towns; Immune system; Instruments; Patient monitoring; Permeability; Voltage control; Electronics, Medical; Humans; Monitoring, Physiologic; Renal Dialysis; Ultrafiltration;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1983.325210
Filename :
4121584
Link To Document :
بازگشت