• DocumentCode
    1211408
  • Title

    An Ultrafiltration Monitor for Hemodialysis Research

  • Author

    Gisser, David ; Strait, Gary ; Zelman, Allen ; Bastidas, Victor ; Harrow, Jeffrey ; Kablitz, Carl ; Stephen, Robert

  • Author_Institution
    Department of Electrical Computer, and Systems Engineering, Rensselaer Polytechnic Institute
  • Issue
    2
  • fYear
    1983
  • Firstpage
    132
  • Lastpage
    136
  • Abstract
    The ultrafitration rate of hemodialysis membranes should be carefully controlled for maximum patient benefit. A special purpose instrument is described that takes in analog pressure information and the analog output of an appropriate weighing system and produces both digital displays and analog outputs (for a strip-chart recorder) of ultrafiltration rate and ultrafiltration index as well as the averaged mean transmembrane hydrostatic pressure drop and volume of ultrafiltrate. The instrument normally uses an averaging time of about 2 min.
  • Keywords
    Artificial biological organs; Biomedical engineering; Biomembranes; Blood; Cities and towns; Immune system; Instruments; Patient monitoring; Permeability; Voltage control; Electronics, Medical; Humans; Monitoring, Physiologic; Renal Dialysis; Ultrafiltration;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1983.325210
  • Filename
    4121584