Title :
Optimization and Design of Permanent Magnet Devices for Retaining Dental Prosthetic Appliances
Author :
Kinouchi, Yohsuke ; Ushita, Tomiyuki ; Tsutsui, Hideo ; Yoshida, Yukiko
Author_Institution :
Department of Electrical Engineering, Faculty of Engineering, Tokushima University
fDate :
4/1/1983 12:00:00 AM
Abstract :
Attraction between a pair of rare earth-cobalt magnets is very useful for retention of dental prosthetic appliances. In these applications, the space available for embedding magnets is very narrow. Therefore, the magnets must have those optimum dimensions which cause the mximum attraction. The optimum dimensions are calculated for a rectangular or disk magnet pair utilizing the axial or radial attraction. In order to maximize the axial attraction of a magnet pair in small separation, the rectangular magnet must be slender and the square or disk magnet must have a size ratio of about OA, which is defined as the ratio of thickness to length of edge or diameter of the magnet. In the case of radial attraction, the optimum disk magnet has a little larger size ratio, e.g., about 0.55 in zero separation. Based on these theoretical analyses, the method of designing the optimum magnets with a required attraction is proposed.
Keywords :
Dentistry; Design methodology; Design optimization; Home appliances; Magnetic analysis; Magnetic materials; Magnetic separation; Magnetization; Permanent magnets; Prosthetics; Cobalt; Dentures; Equipment Design; Magnetics; Mathematics; Metals, Rare Earth;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.1983.325218