• DocumentCode
    1211722
  • Title

    Goid-Doped Silicon Detectors for the Control of Bubble Chamber Photography

  • Author

    Sun, C.R.

  • Author_Institution
    Princeton-Pennsylvania Accelerator Princeton University, Princeton, New Jersey
  • Volume
    9
  • Issue
    3
  • fYear
    1962
  • fDate
    6/1/1962 12:00:00 AM
  • Firstpage
    211
  • Lastpage
    212
  • Abstract
    The properties of a gold-doped silicon detector have been studied with ß-ray sorces. The observed peak of 0.6 Mev from a Sr90-Y90 source agrees well with the most probable energy loss in the detector, calculated for minimum ionizing electrons. Our ultimate aim is to use a matrix of 20 × 34 of such detectors to control our bubble chamber photography.
  • Keywords
    Copper; Electrons; Energy loss; Etching; Gold; Nitrogen; Photography; Pulse amplifiers; Radiation detectors; Silicon;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IRE Transactions on
  • Publisher
    ieee
  • ISSN
    0096-2015
  • Type

    jour

  • DOI
    10.1109/TNS2.1962.4315996
  • Filename
    4315996